Additel ADT227 Documenting Multifunction Process Calibrator

SKU
ADT227
Manufacturer:
Additel
Additel

Key Features:

  • Sourcing, simulating and measuring pressure, temperature and electrical signals
  • Built-in Full Hart Communicator (ADT227-HART)
  • Built-in Barometer
  • Large Smartphone Like Touchscreen User Experience
  • USB Type-C and Bluetooth communications
  • IP67 Rated
  • High Voltage Measurement Capability (±300V AC)
  • True RMS Voltage Meter Capability
  • Dual channel pressure module ports
  • High Static Differential Pressure Measurement 0.002% FS (With Optional Pressure Modules)
Product Options:
$7,255.00

Secured and trusted checkout with

The Additel ADT227 is a portable Multi-function documenting process calibrator. The ADT227 has an easy to use color touchscreen display and provides much improved accuracies. The ADT227 has the option to include the HART communicaton (ADT227-HART).

The ADT227-HART communicator is built-in and will work with most HART transmitters. The ADT227-HART contains an extensive DD library to meet the needs of your smart transmitter.

 

Accessories Included
Model NumberDescriptionQTY
9811-X110V/220V external power adapter (Only for ADT226)1 pc
9704Chargeable Li-ion battery (Only for ADT226)1 pc
9023Test leads1 set (6 pcs)
9027Right angle test leads (Non-ex models only)1 set (2 pcs)
9060Pressure module connection cable1 pc
9052USB Cable type A to type C (Non-Ex models only)1 pc
9040Hanging strap with magnet1 pc
 Manual1 pc

 

Electrical Specifications - Source Accuracy

Specifications

ADT227

ADT227Ex

Range

Resolution

Accuracy

Range

Resolution

Accuracy

Voltage DC

0 to 15 V

0.1 mV

0.005%RDG+0.005%FS

0 to 10.5 V

0.1 mV

0.01%RDG+0.005%FS

Current DC

0 to 25 mA

0.1 uA

0.01%RDG+0.005%FS

0 to 25 mA

0.1 uA

0.01%RDG+0.005%FS

Resistance

0 to 400 Ω

1 mΩ

0.005%RDG+0.005%FS

0 to 400 Ω

1 mΩ

0.01%RDG+0.005%FS

0 to 4000 Ω

10 mΩ

0.01%RDG+0.005%FS

0 to 4000 Ω

10 mΩ

0.01%RDG+0.005%FS

Frequency

0.01 to 50000.0 Hz

Auto range, 6-digit

0.002%RDG+2 on last digit

0.01 to 50000.0 Hz

Auto range, 6-digit

0.002%RDG+2 on last digit

Voltage mV (TC)

-10 to 75 mV

1 uV

0.008%RDG + 0.004%FS

-10 to 75 mV

1 uV

0.01%RDG + 0.004%FS

Pulse

0 to 9999999

1

N/A

0 to 9999999

1

N/A

Optional rising edge and falling edge, minimum threshold voltage: 2.5V

Loop power (max 25mA)

24 V

N/A

±1 V

20 V

N/A

± 10%

 

Electrical Specifications - Measurement Accuracy

Specifications

ADT227

ADT227Ex

Range

Resolution

Accuracy

Range

Resolution

Accuracy

 

 

 

Voltage DC

-300 to 300 mV

1 μV

0.005% RDG + 0.005% FS

-300 to 300 mV

1μV

0.01% RDG + 0.005% FS

-30 to 30 V

0.1 mV

0.005% RDG + 0.005% FS

-30 to 30 V

0.1 mV

0.01% RDG + 0.005% FS

Temperature Coefficiet: ±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C)

Impedance: -300 mV to 300 mV => 100 MΩ, -30 mV to 30 mV => 1 MΩ

 

 

 

 

DC High Voltage

-300 to 300 V

10 mV

 0.05% RDG + 0.01% FS

 

 

 



N/A

Temperature coefficient: ±0.0025% FS/°C (-10°C to 10°C and 30°C to 50°C)

Maximum input voltage = 300 V, IEC61010 300V CATII

Commong mode rejection: >100 dB (at 50 or 60 Hz)

Impedance: > 4 MΩ, DC coupling

 

 

 

 

AC High Voltage

300V (40 to 500 Hz)

10 mV

0.5% RDG + 0.05% FS

 

 

 



N/A

Temperature coefficient: ±(0.025% RD + 0.0025% FS)/°C (-10°C to 10°C and 30°C to 50°C)

Maximum input voltage = 300 V, IEC61010 300V CATII

9% to 100% of range is suitable for the above accuracy indicators

Impedance: >4 MΩ, <100pF, AC coupling

Current DC

-30 to 30 mA

0.1 μA

0.01% RDG + 0.005% FS

-30 to 30 mA

0.1 μA

0.01% RDG + 0.005% FS

Temperature Coefficient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C), Impedance: < 40 Ω

 

 

 

 

Resistance (4-Wire)

0 to 400 Ω

1 mΩ

0.005% RDG + 0.005% FS

0 to 400 Ω

1 mΩ

0.01% RDG + 0.005% FS

0 to 4000 Ω

10 mΩ

0.01% RDG + 0.005% FS

0 to 4000 Ω

10 mΩ

0.01% RDG + 0.005% FS

Temperature coeficient: ±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C)

2-Wire + 50 mΩ, 3-wire+ 10 mΩ

Excitation current: 0.25 mA

 

 

Voltage mV (TC)

-10 to 75 mV

0.1 μV

0.008% RDG + 0.004% FS

-10 to 75 mV

0.1 μV

0.01% RDG + 0.004% FS

Temperature Coefficient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C)

Impedance: >100 MΩ

 

 

Frequency

0.01 to 50000 Hz

Auto range, 6-digit

0.002% RDG + 2 on last digit

0.01 to 50000 Hz

Auto range, 6-digit

0.002% RDG + 2 on last digit

Minimum threshold voltage: 2.5 V

Supported units: Hz, kHz, MHz, CPM, CPH, s, ms, µs

Pulse

0 to 9999999

1

N/A

0 to 9999999

1

N/A

Optional rising edge and falling edge, minimum threshold voltage: 2.5V

Switch

Supports dry or wet switches. Voltage range of 3 to 30 V. Response speed < 10 ms

 

General Specification

 

ADT227

ADT227Ex

Operating Temperature

-10°C to 50°C

Specification guaranteed temperature range

10°C to 30°C

Storage Temperature

-20°C to 60°C

Humidity

<90%, non-condensing

Power supply

6600mAh, 23.8Wh lithium battery, charging time 4~6 hours, battery pack can be charged independently

4000mAh 14.4Wh Explosion-proof lithium battery packcharging time 6~8 hours, battery pack can be charged independently

User interface

Icon drive menus

Icon driven menus with navigation buttons

Ports protection voltage

50V max

30V max

Display

5.0 inch 480 x 800 mm TFT LCD capacitive screen

4.4 inch 640 x 480 mm color display capacitive screen

Maximum altitude

3000 meters

European Compliance

CE Mark

Electrical Connection

Ø4mm sockets and flat mini-jack thermocouple socket

Size

6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm)

Weight

1.6 lb (0.7 kg)

1.65 lb (0.75 kg)

Battery

Rechargeable Li-ion battery (included)

Battery Life

Typically 16 hours

Typically 35 hours

Battery Charge

110V/220V external power adapter included. Battery can be charged external to the unit. Typically charge time is 6-8 hours.

External pressure module

Dual channel serial plug, can connect two digital pressure modules

Warm-up time

Full specification performance is achieved after a 10 minute warm-up time.

ROHS compliant

Rohs II Directive 2011/65/EU, EN50581:2012

Display rate

3 readings per second

Barometric Accuracy (Built-in barometer)

55Pa

IP protection level

IP67, 1 meter drop test

Communication

Isolate USB-TYPEC (slave), Bluetooth BLE

Calibration

ISO 17025 accredited calibration with data

Warranty

3 years

 

Pressure Specification( ADT227 & ADT227Ex)

The ADT161 and ADT161Ex series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar). Accuracy from 0.02% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detailed specifications, please refer to the pressure modules datasheet.

 

Temperature Specification

Thermocouple Measurement and Source Accuracy

 

ADT227

ADT227Ex

Type

Standard

Temperature Range (°C)

Accuracy (°C)

Standard

Temperature Range (°C)

Accuracy (°C)

Measure / Source

Measure / Source

 

 

S

 

 

IEC 584

 

 

-50 to 1768

-50~0

0.76

 

 

IEC 584

 

 

-50 to 1768

-50~100

0.77

0~100

0.56

100~1000

0.42

100~1768

0.44

1000~1768

0.47

 

 

R

 

 

IEC 584

 

 

-50 to 1768

-50~0

0.82

 

 

IEC 584

 

 

-50 to 1768

-50~0

0.82

0~200

0.57

0~200

0.57

200~1768

0.38

200~1768

0.42

 

 


B

 

 


IEC 584

 

 


0 to 1820

200~300

1.51

 

 


IEC 584

 

 


0 to 1820

200~300

1.51

300~500

1

300~500

1

500~800

0.62

500~800

0.62

800~1820

0.43

800~1820

0.43

 

 


K

 

 


IEC 584

 

 


-270 to 1372

-250 to -200

0.72

 

 


IEC 584

 

 


-270 to 1372

-250 to -200

0.75

-200 to -100

0.23

-200 to -100

0.24

-100 to 600

0.12

-100 to 600

0.13

600 to 1372

0.22

600 to 1372

0.25

 

 

N

 

 

IEC 584

 

 

-270 to 1300

-250 to -200

1.14

 

 

IEC 584

 

 

-270 to 1300

-250 to -200

1.17

-200 to -100

0.33

-200 to -100

0.34

-100 to 1300

0.19

-100 to 1300

0.22

 

 


E

 

 


IEC 584

 

 


-270 to 1000

-250~-200

0.39

 

 


IEC 584

 

 


-270 to 1000

-250~-200

0.41

-200~-100

0.15

-200~-100

0.15

-100~700

0.09

-100~700

0.1

700~1000

0.12

700~1000

0.14

 

 

J

 

 

IEC 584

 

 

-210~1200

-210~-100

0.19

 

 

IEC 584

 

 

-210~1200

-210~-100

0.2

-100~700

0.1

-100~700

0.11

700~1200

0.15

700~1200

0.17

 

 

T

 

 

IEC 584

 

 

-270 to 400

-250~-100

0.55

 

 

IEC 584

 

 

-270 to 400

-250~-100

0.57

-100~0

0.12

-100~0

0.23

0~400

0.08

0~400

0.08

 

 

C

 

 

ASTM E988

 

 

0 to 2315

0 to 1000

0.24

 

 

ASTM E988

 

 

0 to 2315

0 to 1000

0.26

1000 to 1800

0.4

1000 to 1800

0.45

1800 to 2315

0.65

1800 to 2315

0.73

 

 


D

 

 


ASTM E988

 

 


0~2315

0~100

0.31

 

 


ASTM E988

 

 


0~2315

0~100

0.31

100~1200

0.25

100~1200

0.27

1200~2000

0.42

1200~2000

0.47

2000~2315

0.65

2000~2315

0.74

 

 

 

G

 

 

 

ASTM E1751

 

 

 

0 to 2315

50~100

0.9

 

 

 

ASTM E1751

 

 

 

0 to 2315

50~100

0.9

100~200

0.57

100~200

0.57

200~400

0.35

200~400

0.36

400~1500

0.25

400~1500

0.27

1500~2315

0.49

1500~2315

0.55

 

 

L

 

 

DIN43710

 

 

-200 to 900

-200 to -100

0.11

 

 

DIN43710

 

 

-200 to 900

-200 to -100

0.12

-100 to 400

0.08

-100 to 400

0.09

400 to 900

0.1

400 to 900

0.12

 

U

 

DIN43710

 

-200 to 600

-200 to 0

0.21

 

DIN43710

 

-200 to 600

-200 to 0

0.21

0 to 600

0.08

0 to 600

0.09

Note: Internal CJC is ±0.15°C (-10°C to 50°C ambient temperature)

Accuracy with external cold junction only, for internal cold junction add 0.15 ℃ (k=2)

 

RTD Measurement and Source Accuracy

Measure and Simulate

Temperature Range (°C)

Accuracy (°C)

ADT227

ADT227Ex

 

 

PT10(385)

 

 

-200 to 850

-200~200

0.57

0.59

200~600

0.67

0.72

600~850

0.75

0.82

 

 

PT25(385)

 

 

-200 to 850

-200~200

0.24

0.27

200~600

0.3

0.35

600~850

0.34

0.41

 

 

PT50(3916)

 

 

-200 to 850

-200~200

0.13

0.16

200~600

0.17

0.22

600~850

0.2

0.27

 

 

PT100(385) PT100(391) PT100(3916) PT100(3926)

 

 

-200 to 850

-200~200

0.08

0.1

200~600

0.11

0.16

600~850

0.14

0.2

 

 

PT200(385)

 

 

-200 to 850

-200~200

0.32

0.08

200~300

0.34

0.34

300~600

0.41

0.41

600~850

0.48

0.48

 

 

PT400(385)

 

 

-200 to 850

-200~0

0.15

0.04

0~200

0.18

0.18

200~600

0.25

0.25

600~850

0.3

0.3

 

 

PT500(385)

 

 

-200 to 850

-200~200

0.16

0.16

200~600

0.22

0.22

600~850

0.27

0.27

 

 

PT1000(385)

 

 

-200 to 850

-200~200

0.1

0.1

200~600

0.16

0.16

600~850

0.2

0.2

Cu10(427)

-200~260

-200~260

0.54

0.56

Cu50(428)

--200~260

-200~260

0.11

0.13

Cu100(428)

-200~260

-200~260

0.07

0.08

Ni100(617) Ni100(618)

-60~180

-60~0

0.05

0.06

0~180

0.05

0.05

Ni120(672)

--80~260

-80~260

0.04

0.05

Ni1000

-50~150

-50~150

0.07

0.07

 