We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Additel ADT226 is a portable multi-functional process calibrator with an easy to use color touchscreen display. The ADT226 is powerful, cost effective, and accurate.
Â
Accessories Included | ||
Model Number | Description | QTY |
9811-X | 110V/220V external power adapter (Only for ADT226) | 1 pc |
9704 | Chargeable Li-ion battery (Only for ADT226) | 1 pc |
9023 | Test leads | 1 set (6 pcs) |
9027 | Right angle test leads kit (Non-Ex models only) | 1 set (2 pcs) |
9060 | Pressure module connection cable | 1 pc |
9052 | USB Cable type A to type C (Non-Ex models only) | 1 pc |
9040 | Hanging strap with magnet | 1 pc |
Â
Electrical Specifications - Source Accuracy | ||||||
Specifications | ADT226 | ADT226Ex | ||||
Range | Resolution | Accuracy | Range | Resolution | Accuracy | |
Voltage DC | 0 to 15 V | 0.1 mV | 0.015%RDG+0.005%FS | 0 to 10.5 V | 0.1 mV | 0.02%RDG+0.005%FS |
Current DC | 0 to 25 mA | 0.1 uA | 0.015%RDG+0.005%FS | 0 to 25 mA | 0.1 uA | 0.02%RDG+0.005%FS |
Resistance | 0 to 400 Ω | 1 mΩ | 0.015%RDG+0.005%FS | 0 to 400 Ω | 1 mΩ | 0.02%RDG+0.005%FS |
0 to 4000 Ω | 10 mΩ | 0.015%RDG+0.005%FS | 0 to 4000 Ω | 10 mΩ | 0.02%RDG+0.005%FS | |
Frequency | 0.01 to 50000.0 Hz | Auto range, 6-digit | 0.005%RDG+5 on last digit | 0.01 to 50000.0 Hz | Auto range, 6-digit | 0.005%RDG+5 on last digit6-digit 0.005%RDG+5 on last digit |
Voltage mV (TC) | -10 to 75 mV | 1 uV | 0.015%RDG + 0.005%FS | -10 to 75 mV | 1 uV | 0.02%RDG+0.005%FS |
Pulse | 0 to 9999999 | 1 | N/A | 0 to 9999999 | 1 | N/A |
Optional rising edge and falling edge, minimum threshold voltage: 2.5V | ||||||
Loop power (max 25mA) | 24 V | N/A | ±1 V | 20 V | N/A | ± 10% |
Â
Electrical Specifications - Measurement Accuracy | ||||||
Specifications | ADT226 | ADT226Ex | ||||
Range | Resolution | Accuracy | Range | Resolution | Accuracy | |
   Voltage DC | -300 to 300 mV | 1 μV | 0.015% RDG + 0.005% FS | -300 to 300 mV | 1μV | 0.02% RDG + 0.005% FS |
-30 to 30 V | 0.1 mV | 0.015%RDG+0.005%FS | -30 to 30 V | 0.1 mV | 0.02% RDG + 0.005% FS | |
Temperature Coefficient: ±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C) | ||||||
Impedance: -300 mV to 300 mV = > 100 MΩ, -30 V to 30 V = >1 MΩ | ||||||
    DC High Voltage | -300 to 300 V | 10 mV | 0.05% RDG + 0.01% FS |  Â
| ||
Temperature coefficient: ±0.0025% FS/°CDC High Voltage N/A(-10°C to 10°C and 30°C to 50°C) | ||||||
The highest input voltage is 300 V, IEC61010 300V CATII | ||||||
Common mode rejection: >100 dB (at 50 or 60 Hz) | ||||||
Impedance: > 4 MΩ, DC coupling | ||||||
    AC High Voltage | 300V (40 to 500 Hz) | 10 mV | 0.5% RDG + 0.05% FS |  Â
| ||
Temperature coefficient: ±(0.025% RD + 0.0025% FS /°C, -10°C to 10°C and 30°C to 50°C) | ||||||
The highest input voltage is 300 V, IEC61010 300V CATII | ||||||
9% to 100% of range is suitable for the above accuracy indicators | ||||||
Impedance: >4 MΩ, <100pF, AC coupling | ||||||
Current DC | -30 to 30 mA | 0.1 μA | 0.015% RDG + 0.005% FS | -30 to 30 mA | 0.1 μA | 0.02% RDG + 0.005% FS |
Temperature Coefficient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C), Impedance: < 40 Ω | ||||||
    Resistance (4-Wire) | 0 to 400 Ω | 1 mΩ | 0.015% RDG + 0.005% FS | 0 to 400 Ω | 1 mΩ | 0.02% RDG + 0.005% FS |
0 to 4000 Ω | 10 mΩ | 0.015% RDG + 0.005% FS | 0 to 4000 Ω | 10 mΩ | 0.02% RDG + 0.005% FS | |
Temperature coeficient: ±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C) | ||||||
2-Wire + 50 mΩ, 3-wire+ 10 mΩ | ||||||
Excitation current: 0.25 mA | ||||||
  Voltage mV (TC) | -10 to 75 mV | 0.1 μV | 0.015% RDG + 0.005% FS | -10 to 75 mV | 0.1 μV | 0.02% RDG + 0.005% FS |
Temperature Coefficient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C) | ||||||
Impedance: >100 MΩ | ||||||
  Frequency | 0.01 to 50000 Hz | Auto range, 6-digit | 0.005% RDG + 5 on last digit | 0.01 to 50000 Hz | Auto range, 6-digit | 0.005% RDG + 5 on last digit |
Minimum threshold voltage: 2.5 V | ||||||
Supported units: Hz, kHz, MHz, CPM, CPH, s, ms, µs | ||||||
Pulse | 0 to 9999999 | 1 | N/A | 0 to 9999999 | 1 | N/A |
Optional rising edge and falling edge, minimum threshold voltage: 2.5V | ||||||
Switch | Support for dry or wet switch, voltage range of 3 to 30 V, response speed of < 10 ms |
Â
General Specification | ||
 | ADT226 | ADT226Ex |
Operating Temperature | -10°C to 50°C | |
Specification guaranteed temperature range | 10°C to 30°C | |
Storage Temperature | -20°C to 70°C | |
Humidity | <95%, non-condensing | |
Power supply | 6600mAh, 23.8Wh lithium battery, charging time 4~6 hours, battery pack can be charged independently | 4000mAh 14.4Wh Explosion-proof lithium battery packcharging time 6~8 hours, battery pack can be charged independently |
User interface | Icon drive menus | Icon driven menus with navigation buttons |
Ports protection voltage | 50V max | 30V max |
Display | 5.0 inch 480 x 800 mm TFT LCD capacitive screen | 4.4 inch 640 x 480 mm color display capacitive screen |
Maximum altitude | 3000 meters | |
European Compliance | CE Mark | |
Electrical Connection | Ø4mm sockets and flat mini-jack thermocouple socket | |
Size | 6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm) | |
Weight | 1.6 lb (0.7 kg) | 1.65 lb (0.75 kg) |
Battery | Rechargeable Li-ion battery (included) | |
Battery Life | Typically 16 hours | Typically 35 hours |
Battery Charge | 110V/220V external power adapter included. Battery can be charged external to the unit. Typically charge time is 6-8 hours. | |
External pressure module | Dual channel serial plug, can connect two digital pressure modules | |
Warm-up time | Full specification performance is achieved after a 10 minute warm-up time. | |
ROHS compliant | Rohs II Directive 2011/65/EU, EN50581:2012 | |
Display rate | 3 readings per second | |
Barometric Accuracy (Built-in barometer) | 55Pa | |
IP protection level | IP67, 1 meter drop test | |
Communication | Isolate USB-TYPEC (slave), Bluetooth BLE | |
Calibration | ISO 17025 accredited calibration with data | |
Warranty | 3 years |
Â
Pressure Specification( ADT226 & ADT226Ex) |
The 161 series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar). Accuracy from 0.02% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detailed specifications, please refer to the pressure modules datasheet. |
Â
Temperature Specification | ||||||||
Thermocouple Measurement and Source Accuracy | ||||||||
 | ADT226 | ADT226Ex | ||||||
Type | Standard | Temperature Range (°C) | Accuracy (°C) | Standard | Temperature Range (°C) | Accuracy (°C) | ||
Measure / Source | Measure / Source | |||||||
  S |   IEC 584 |   -50 to 1768 | -50~0 | 0.96 |   IEC 584 |   -50 to 1768 | -50~100 | 0.96 |
0~100 | 0.69 | 100~1000 | 0.69 | |||||
100~1768 | 0.64 | 1000~1768 | 0.73 | |||||
  R |   IEC 584 |   -50 to 1768 | -50~0 | 1.02 |   IEC 584 |   -50 to 1768 | -50~0 | 1.03 |
0~200 | 0.71 | 0~200 | 0.71 | |||||
200~1768 | 0.56 | 200~1768 | 0.65 | |||||
  B |   IEC 584 |   0 to 1820 | 200~300 | 1.89 |   IEC 584 |   0 to 1820 | 200~300 | 1.9 |
300~500 | 1.25 | 300~500 | 1.26 | |||||
500~800 | 0.78 | 500~800 | 0.79 | |||||
800~1820 | 0.55 | 800~1820 | 0.57 | |||||
  K |   IEC 584 |   -270 to 1372 | -250 to -200 | 0.97 |   IEC 584 |   -270 to 1372 | -250 to -200 | 1.04 |
-200 to -100 | 0.3 | -200 to -100 | 0.32 | |||||
-100 to 600 | 0.18 | -100 to 600 | 0.21 | |||||
600 to 1372 | 0.35 | 600 to 1372 | 0.43 | |||||
  N |   IEC 584 |   -270 to 1300 | -250 to -200 | 1.5 |   IEC 584 |   -270 to 1300 | -250 to -200 | 1.58 |
-200 to -100 | 0.44 | -200 to -100 | 0.46 | |||||
-100 to 1300 | 0.3 | -100 to 1300 | 0.37 | |||||
  E |   IEC 584 |   -270 to 1000 | -250~-200 | 0.54 |   IEC 584 |   -270 to 1000 | -250~-200 | 0.59 |
-200~-100 | 0.2 | -200~-100 | 0.22 | |||||
-100~700 | 0.15 | -100~700 | 0.18 | |||||
700~1000 | 0.2 | 700~1000 | 0.25 | |||||
  J |   IEC 584 |   -210~1200 | -210~-100 | 0.26 |   IEC 584 |   -210~1200 | -210~-100 | 0.28 |
-100~700 | 0.15 | -100~700 | 0.19 | |||||
700~1200 | 0.25 | 700~1200 | 0.31 | |||||
  T |   IEC 584 |   -270 to 400 | -250~-100 | 0.74 |   IEC 584 |   -270 to 400 | -250~-100 | 0.79 |
-100~0 | 0.15 | -100~0 | 0.16 | |||||
0~400 | 0.11 | 0~400 | 0.13 | |||||
  C |   ASTM E988 |   0 to 2315 | 0 to 1000 | 0.35 |   ASTM E988 |   0 to 2315 | 0 to 1000 | 0.4 |
1000 to 1800 | 0.62 | 1000 to 1800 | 0.73 | |||||
1800 to 2315 | 1.02 | 1800 to 2315 | 1.22 | |||||
   D |    ASTM E988 |    0~2315 | 0~100 | 0.39 |    ASTM E988 |    0~2315 | 0~100 | 0.39 |
100~1200 | 0.37 | 100~1200 | 0.43 | |||||
1200~2000 | 0.65 | 1200~2000 | 0.77 | |||||
2000~2315 | 1.03 | 2000~2315 | 1.24 | |||||
   G |    ASTM E1751 |    0 to 2315 | 50~100 | 1.12 |    ASTM E1751 |    0 to 2315 | 50~100 | 1.12 |
100~200 | 0.72 | 100~200 | 0.72 | |||||
200~400 | 0.45 | 200~400 | 0.46 | |||||
400~1500 | 0.37 | 400~1500 | 0.43 | |||||
1500~2315 | 0.77 | 1500~2315 | 0.92 | |||||
  L |   DIN43710 |   -200 to 900 | -200 to -100 | 0.15 |   DIN43710 |   -200 to 900 | -200 to -100 | 0.16 |
-100 to 400 | 0.13 | -100 to 400 | 0.14 | |||||
400 to 900 | 0.17 | 400 to 900 | 0.2 | |||||
 U |  DIN43710 |  -200 to 600 | -200 to 0 | 0.28 |  DIN43710 |  -200 to 600 | -200 to 0 | 0.29 |
0 to 600 | 0.13 | 0 to 600 | 0.15 | |||||
Note: Internal CJC is ±0.15°C (-10°C to 50°C ambient temperature) | ||||||||
Accuracy with external cold junction only, for internal cold junction add 0.15 ℃ (k=2) |
Â
RTD Measurement and Source Accuracy | ||||
Measure and Simulate | Temperature Range (°C) | Accuracy (°C) | ||
ADT226 | ADT226Ex | |||
  PT10(385) |   -200 to 850 | -200~200 | 0.62 | 0.64 |
200~600 | 0.77 | 0.82 | ||
600~850 | 0.88 | 0.95 | ||
  PT25(385) |   -200 to 850 | -200~200 | 0.29 | 0.31 |
200~600 | 0.4 | 0.44 | ||
600~850 | 0.47 | 0.54 | ||
  PT50(3916) |   -200 to 850 | -200~200 | 0.18 | 0.2 |
200~600 | 0.27 | 0.32 | ||
600~850 | 0.34 | 0.4 | ||
  PT100(385) PT100(391) PT100(3916) PT100(3926) |   -200 to 850 | -200~200 | 0.13 | 0.15 |
200~600 | 0.21 | 0.26 | ||
600~850 | 0.27 | 0.34 | ||
   PT200(385) |    -200 to 850 | -200~200 | 0.34 | 0.12 |
200~300 | 0.37 | 0.4 | ||
300~600 | 0.46 | 0.51 | ||
600~850 | 0.54 | 0.61 | ||
   PT400(385) |    -200 to 850 | -200~0 | 0.17 | 0.06 |
0~200 | 0.21 | 0.23 | ||
200~600 | 0.3 | 0.35 | ||
600~850 | 0.37 | 0.44 | ||
  PT500(385) |   -200 to 850 | -200~200 | 0.18 | 0.2 |
200~600 | 0.27 | 0.32 | ||
600~850 | 0.34 | 0.4 | ||
  PT1000(385) |   -200 to 850 | -200~200 | 0.13 | 0.15 |
200~600 | 0.21 | 0.26 | ||
600~850 | 0.27 | 0.34 | ||
Cu10(427) | -200~260 | -200~260 | 0.59 | 0.61 |
Cu50(428) | --200~260 | -200~260 | 0.15 | 0.17 |
Cu100(428) | -200~260 | -200~260 | 0.1 | 0.12 |
Ni100(617) Ni100(618) | -60~180 | -60~0 | 0.06 | 0.07 |
0~180 | 0.06 | 0.08 | ||
Ni120(672) | --80~260 | -80~260 | 0.06 | 0.07 |
Ni1000 | -50~150 | -50~150 | 0.08 | 0.09 |
Â