Newport 201AN & 201A Low-Cost 3-1/2 Digit DC Voltmeters
|
Item Number: 201
Manufacturer: Newport Electronics
Part Number: 201
Product Options
* denotes required field
Newport 201 Model Selection*
Signal Input Range Selection*
3-Wire Ratio Input Impedance Selection*
Decimal Point Selection*
Power Selection*
Display LEDs Color Selection*
Screw-Terminal Barrier Strip Selection (201AN Model Only)*
Newport Logo Option*
Newport NEMA 4 Splash Proof Meter Cover*
Panel Meter NIST Calibration & Scaling*
If Ordering Calibration and Scaling, Please Specify Input Signal and Scale
|
|
Newport 201AN & 201A Low-Cost 3-1/2 Digit DC Voltmeters
Features:
- +/-1,999-count display span
- +/-200 mV to +/-200 Vdc ranges
- True-differential input with 80 dB CMR
- 3-wire ratio input
- Bright, 0.56 in (14.2 mm) LED display
- Automatic zero and polarity
- Display hold and test
- 115/230 Vac power
- 1/8 DIN case
Options:
- Isolated 9-32 Vdc power
- Isolated 26-56 Vdc power
- Screw-terminal barrier strip
- NEMA-4 splash-proof lens cover
Models 201A and 201AN Are two pinout versions of the same low-cost 3 1/2 digit DC voltmeter. They fit a standard 1/8 DIN panel cutout and require a depth of less than 4.1 in (104 mm) behind the panel. With the addition of a screw-terminal barrier strip, they become low-cost, pin-compatible alternatives to the Q2000A and Q2000B in applications where indication-only is required.
Input Ranges: The 201A and 201AN offer voltage ranges of +/-0.1999 V, +/-1.999 V, +/-19.99 V and +/-199.9 V dc. The input is true-differential, with excellent common-mode noise rejection (CMR). By using an external DC voltage reference, any of these ranges can further be configured for 3-wire ratio measurement with readout from 0 to 1.999.
Two Pinouts: Model 201A is pin-compatible with Newport's older Model 201 and is recommended for existing 201 applications. Model 201AN is pin-compatible with Newport's new 202A, 2001A, 2002A Series and is recommended for new applications.
Options: The 201A and 201AN can be configured for four voltage input ranges, three decimal-point positions, and high- impedance 3-wire ratio input. Mechanical options include a screw-terminal barrier strip for signal and power and a splash-proof lens cover which meets NEMA-4 standards.
Specifications:
Analog Input
|
Range |
+/-199.9 V |
+/-1.999 V |
+/-19.99 V |
+/-199.9 V |
|
Resolution |
100 uV |
1 mV |
10 mV |
100 mV |
|
Maximum Input |
130 Vp |
130 Vp |
250 Vp |
250 Vp |
|
Input Resistance |
1 Gohm |
1 Gohm |
1.1 Mohm |
1.0 Mohm |
|
Bias Current |
50 pA |
50 pA |
5 pA |
1 pA |
Configuration: Differential, Bipolar Zero: Automatic Span Adjustment: +/-5%
Noise Rejection: NMR: 56 dB, 50/60 Hz CMR, AC GND to SIG GND: 120 dB, DC to 60 Hz CMR, SIG LO to SIG GND: 80 dB, DC to 60 Hz CMV, AC GND to SIG GND: 1500 Vp per HV test, 354 Vp per IEC spacing CMV, SIG LO to SIG GND: +/-1 Vp
Accuracy at 25 C: Maximum error: +/-0.05% of reading +/-1 count Span tempco: +/-0.01% of reading/Degree C Step response: 1 s Warmup to rated accuracy: 1 min
Analog-to-Digital Conversion: Technique: Dual-slope, average-value Polarity: Automatic Signal integration period: 100 ms
Option SPC4 NEMA 4 Splash Proof Meter Cover:

|
|


|
|