Fluke 80i-110s AC/DC Current Clamp
|
|
Item Number: 80i-100s
Manufacturer: Fluke
Part Number: 80i-100s
|
Fluke 80i-110s AC/DC Current Clamp
The Fluke 80I-110s is compatible with any Fluke ScopeMeter Test Tool, Power Harmonics Analyzer, Oscilloscope, Multimeter, or other voltage measurement device that has the following features: BNC input connectors or if using a meter convert the banana plugs to a BNC connector using the adapter PM9081/001.
Features:
-
Current Range:0.1 to 100A DC or 0.1 to 70A AC
-
Basic Accuracy (DC to 1kHz):
100mV/A setting:50mA to 10A +/- 3% of reading + 50mA
10mV/A setting: 50mA to 40A +/- 4% of reading + 50mA
40A to 80A +/- 12% of reading + 50mA
80 to 100A +/- 15% of reading
-
Output Signal:10A range: 100mV/A | 100A range: 10mV/A
-
Bandwidth:1Hz to 20kHz
-
Working Voltage:600V AC rms
-
Maximum Conductor Size:11.8 mm (.46 inch) diameter
-
Safety Rating:IEC 1010 and CSA-C22.2 No. 1010-1 Installation
Category II, Working Voltage 600V
Specifications:
|
Nominal Current Range
|
10 A, AC/DC
100 A, AC/DC
|
|
Continuous Current Range
|
0.1 A - 100 A AC/DC
1 A - 100 A AC/DC
|
|
Maximum Non-Destructive Current
|
140 A - 2 kHz
|
|
Lowest Measurable Current
|
0.1 A
|
|
Basic Accuracy
|
3% + 50 mA (@ 10A)
(% reading + floorspec)
|
|
Useable Frequency
|
DC - 100 kHz
|
|
Output Level(s)
|
100 mV/A
10 mV/A
|
|
Zero Error Adjustment
|
Yes
|
Safety Specifications:
|
Safety
|
CAT II, 600 V
CAT III, 300 V
|
|
Miximum Voltage
|
600 V
|
Mechanical & General Specifications:
| Warranty
|
1 Year
|
| Battery Life
|
9 V, 55 h
|
| Maximum Conductor Diameter
|
11.8 mm
|
| Output Cable Length
|
1.6 m
|
| BNC Adapter
|
Yes
|
|
|


|
|