Fluke 753 Documenting Process Calibrator
|
|
Item Number: Fluke-753
Manufacturer: Fluke
Part Number: Fluke-753
Product Options
Fluke 750 Series Calibration Options
|
Fluke 753 Documenting Process Calibrator

Fans of the time tested Fluke 743B have a reason to rejoice! Fluke presents the new and better Fluke 753 as a direct replacement for the ever popular Fluke 743. There is no need to retrain technicians, the new user interface is nearly identical to the popular 740 series calibrators and an improved display makes these new calibrators a winner for those used to using 740 series calibrators.
Improvements include a new brighter, bigger and easier to read display, HART (Fluke 754 only) connectability through the mA terminals, improved test lead set, Li-ion battery, improved RTD accuracy and more. Added bonus, the new 750 series DPCs include the innovative new C799 Field Softcase.
The Fluke 753 Documenting Process Calibrator automates calibration procedures and captures your data. Use the PC interface for downloading procedures, lists, and instructions to the 753 and for uploading data for printing, archiving and analysis.

Features of the Fluke 753 and 754 Process Calibrators:
• Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
• Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
• Power transmitters during test using loop supply with simultaneous mA measurement
• Measure/source pressure using any of 29 Fluke pressure modules
• Measure and source simultaneously with one compact, rugged, reliable tool
• Create and run automated as-found/as-left procedures to satisfy quality programs or regulations. Record and document results
• Advanced features like autostep, custom units, user entered values during test, one-point and two-point switch testing,
square root DP flow testing, programmable measurement delay and more
• Control the temperature of selected Hart Scientific Dry Blocks
• Handle fast pulsed RTD transmitters and PLCs in RTD source mode with 1 ms response time
• English, French, German, Italian, and Spanish languages
• Three-year warranty
Fluke 754 Documenting Process Calibrator is more advanced than the Fluke 753 by adding:
The 754 includes integrated HART communication capability. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART and other instrumentation.

Additional 754 features:
• Monitor, control and calibrate HART instrumentation
• Interrogate HART devices to determine type, manufacturer,model and tag-ID
• Read HART PV, smart transmitter digital output
• Make field adjustments to ranging, damping and more
• Change/assign the HART transmitter tag
• Re-configure HART temperature sensor (e.g., TC to RTD)
• Perform HART sensor trim and output trim
• Perform HART loop test
FLUKE-753 Documenting Process Calibrator, FLUKE-754 Documenting Process Calibrator-HART include:

Specifications - Fluke 753 & Fluke 754 Process Calibrators
| Measurement
Accuracy
|
| Voltage DC
|
|
1
Year
|
2
Years
|
| 100.000 mV
|
0.02%+0.005% mV
|
0.03%+0.005 mV
|
| 3.00000 V
|
0.02%+0.00005% V
|
0.03%+0.00005 V
|
| 30.000 V
|
0.02%+0.0005% V
|
0.03%+0.0005 V
|
| 300.00 V
|
0.05%+0.05% V
|
0.07%+0.05 V
|
|
| Voltage AC
|
Range
40 to 500 Hz
|
Resolution
|
1
Year
|
2
Years
|
| 3.000 V
|
0.001
V
|
0.5%+0.002 V
|
1.0%+0.004 V
|
| 30.00 V
|
0.01
V
|
0.5%+0.02 V
|
1.0%+0.04 V
|
| 300.0 V
|
0.1
V
|
0.5%+0.2
V
|
1.0%+0.2
V
|
|
| Current DC
|
|
|
1
Year
|
2
Years
|
| 30.000 mA
|
0.01%
+ 5 uA
|
0.015%+7
uA
|
| 110.00 mA
|
0.01% +
20 uA
|
0.015%+30 uA
|
|
| Resistance
|
|
|
1
Year
|
2
Years
|
10.000 O
|
0.05% +
50 mO
|
0.07%+70
mO
|
| 100.00 O
|
0.05% +
50 mO
|
0.07%+70
mO
|
| 1.0000 kO
|
0.05% +
500 mO
|
0.07%+0.5 O
|
| 10.000 kO
|
0.1% +
10 O
|
0.15%+15
O
|
|
| Frequency
|
|
|
Resolution
|
Accuracy (2
Years)
|
| 1.00 to 110.00 Hz
|
0.01
Hz
|
0.05
Hz
|
| 110.1 to 1100.0 Hz
|
0.1
Hz
|
0.5
Hz
|
| 1.101 to 11.000 kHz
|
0.001
kHz
|
0.005
kHz
|
| 11.01 to 50.00 kHz
|
0.01
kHz
|
0.05
kHz
|
|
| Source
Accuracy
|
| Voltage DC
|
|
|
1
Year
|
2 Years
|
100.000 mV
|
0.01%+0.005% mV
|
0.015%+0.005% mV
|
| 1.00000 V
|
0.01%+0.00005% V
|
0.015%+0.0005% V
|
| 15.0000 V
|
0.01%+0.0005% V
|
0.015%+0.0005%
V
|
|
| Current DC
|
|
|
1
Year
|
2 Years
|
| 22.000 mA (Source)
|
0.01%+
0.003 mA
|
0.02%+
0.003 mA
|
| Current Sink (Simulate)
|
0.02% +
0.007 mA
|
0.04% +
0.007 mA
|
|
| Resistance
|
|
|
1
Year
|
2
Years
|
10.000 O
|
0.01% +
10 mO
|
0.015% +
15 mO
|
| 100.00 O
|
0.01% +
20 mO
|
0.015% +
30 mO
|
| 1.0000 kO
|
0.02%
+ 0.2 O
|
0.03%
+ 0.3 O
|
| 10.000 kO
|
0.02% +
3 O
|
0.03% +
5 O
|
|
| Frequency
|
|
|
2
Years
|
| 0.1 to 10.99 Hz
|
0.01
Hz
|
| 0.01 to 10.99 Hz
|
0.01
Hz
|
| 11.00 to 109.99 Hz
|
0.1
Hz
|
| 110.0 to 1099.9 Hz
|
0.1
Hz
|
| 1.100 to 21.999 kHz
|
0.002
kHz
|
22.000 to 50.000 kHz
|
0.005
kHz
|
|
| Technical
Data
|
| Data Log Functions
|
| Measure functions:
|
Voltage,
current, resistance, frequency, temperature, pressure
|
| Reading rate:
|
1, 2, 5,
10, 20, 30, or 60 readings/minute
|
| Maximum record length:
|
8000
readings (7980 for 30 or 60
readings/minute)
|
|
| Ramp Functions
|
| Source functions:
|
Voltage,
current, resistance, frequency, temperature
|
| Rate:
|
4
steps/second
|
| Trip detect:
|
Continuity or voltage (continuity detection not available
when sourcing current)
|
|
| Loop Power Function
|
| Voltage:
|
Selectable, 26 V
|
| Accuracy:
|
10%, 18
V minimum at 22 mA
|
| Maximum current:
|
25 mA,
short circuit protected
|
| Maximum input voltage:
|
50 V
DC
|
|
| Step Functions
|
| Source Functions
|
Voltage,
current, resistance, frequency, temperature
|
| Manual Step
|
Selectable step, change with arrow
buttons
|
| Autostep
|
Fully
programmable for function, start delay, stepvalue, time per step,
repeat
|
|
| Environmental
Specifications
|
| Operating Temperature
|
|
| Storage Temperature
|
|
| Dust/water resistance
|
|
| Operating Altitude
|
|
3000 m
above mean sea level (9842
ft)
|
|
| Safety
Specifications
|
| Agency Approvals
|
|
CAN/CSA
C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and
EN610-1:1993
|
|
| Mechanical & General
Specifications
|
| Size
|
|
136 x
245 x 63 mm (5.4“ x 9.6 x 2.5)
|
|
| Weight
|
|
| Batteries
|
| Internal Battery Pack Li-ion:
|
7.2V, 4400mAh, 30
Wh
|
|
| Battery Life
|
|
| Battery Replacement
|
|
Via
snap-shut door without opening calibrator; no tools
required
|
|
| Side port connections
|
|
Pressure
module connector
|
|
USB
Connector to interface to your PC
|
|
|
Digital
instrument (HART) connector
|
|
Connection for optional battery charger/eliminator
|
|
| Data storage capacity
|
|
1 week
of calibration procedures results
|
|
| 90 day specifications
|
|
The
standard specification interval for the 750 Series are 1 and 2
years.
|
|
Typical
90 day measurement and source accuracy can be estimated by
|
|
dividing
the one year "% of reading" or "%of output" specifications by
2.
|
|
Floor
specifications, expressed as "% of full scale" or
|
|
"counts"
or "ohms" remain
constant.
|
|
| Temperature,
Resistance Temperature DetectorsHBZSEL28
|
| Degrees or % of reading
|
| Type
(a)
|
Range
°C
|
Measure °C 2
|
Source
current
|
Source °C
|
Allowable current
3
|
|
|
1
year
|
2
year
|
|
1
year
|
2
year
|
|
| 100 O Pt (385)
|
-200 to 100
100 to 800
|
0.07 °C
0.02 % + 0.05
°C
|
0.14 °C
0.04 % + 0.10
°C
|
1 mA
|
0.05 °C
0.0125 % + 0.04
°C
|
0.10 °C
0.025 % + 0.08
°C
|
0.1 mA to 10 mA
|
| 200 O Pt (385)
|
-200 to 100
100 to 630
|
0.07 °C
0.02 % + 0.05
°C
|
0.14 °C
0.04 % + 0.10
°C
|
500 µA
|
0.06 °C
0.017 % + 0.05
°C
|
0.12 °C
0.034 % + 0.10
°C
|
0.1 mA to 1 mA
|
| 500 O Pt (385)
|
-200 to 100
100 to 630
|
0.07 °C
0.02 % + 0.05
°C
|
0.14 °C
0.04 % + 0.10
°C
|
250 µA
|
0.06 °C
0.017% + 0.05
°C
|
0.12 °C
0.034 % + 0.10
°C
|
0.1 mA to 1 mA
|
| 1000 O Pt (385)
|
-200 to 100
100 to 630
|
0.07 °C
0.02 % + 0.05
°C
|
0.14 °C
0.04% + 0.10 °C
|
150 µA
|
0.06 C
0.017 % + 0.05
°C
|
0.12 C
0.034 % + 0.10
°C
|
0.1 mA to 1 mA
|
| 100 O Pt (3916)
|
-200 to 100
100 to 630
|
0.07 °C
0.02 % +0.05 °C
|
0.14 °C
0.04 % +0.10 °C
|
1 mA
|
0.05 °C
0.0125 % + 0.04
°C
|
0.10 °C
0.025 % + 0.08
°C
|
0.1 mA to 10 mA
|
| 100 O Pt (3926)
|
-200 to 100
100 to 630
|
0.08 °C
0.02 % +0.06 °C
|
0.16 °C
0.04 % +0.12 °C
|
1 mA
|
0.05 °C
0.0125 % + 0.04
°C
|
0.10 °C
0.025 % + 0.08
°C
|
0.1 mA to 10 mA
|
| 10 O Cu (427)
|
-100 to 260
|
0.2 °C
|
0.4 °C
|
3 mA
|
0.2 °C
|
0.4 °C
|
0.1 mA to 10 mA
|
| 120 O Ni (672)
|
-80 to 260
|
0.1 °C
|
0.2 °C
|
1 mA
|
0.04 °C
|
0.08 °C
|
0.1 mA to 10 mA
|
| Temperature, Thermocouples
|
| Type
|
Source °C
|
Measure °C
|
Source °C
|
|
|
1 year
|
2 years
|
1 year
|
2 years
|
|
E
|
-250 to -200
|
1.3
|
2.0
|
0.6
|
0.9
|
|
-200 to -100
|
0.5
|
0.8
|
0.3
|
0.4
|
|
-100 to 600
|
0.3
|
0.4
|
0.3
|
0.4
|
|
600 to 1000
|
0.4
|
0.6
|
0.2
|
0.3
|
|
N
|
-200 to -100
|
1.0
|
1.5
|
0.6
|
0.9
|
|
-100 to 900
|
0.5
|
0.8
|
0.5
|
0.8
|
|
900 to 1300
|
0.6
|
0.9
|
0.3
|
0.4
|
|
J
|
-210 to -100
|
0.6
|
0.9
|
0.3
|
0.4
|
|
-100 to 800
|
0.3
|
0.4
|
0.2
|
0.3
|
|
800 to 1200
|
0.5
|
0.8
|
0.3
|
0.3
|
|
K
|
-200 to -100
|
0.7
|
1.0
|
0.4
|
0.6
|
|
-100 to 400
|
0.3
|
0.4
|
0.3
|
0.4
|
|
400 to 1200
|
0.5
|
0.8
|
0.3
|
0.4
|
|
1200 to 1372
|
0.7
|
1.0
|
0.3
|
0.4
|
|
T
|
-250 to -200
|
1.7
|
2.5
|
0.9
|
1.4
|
|
-200 to 0
|
0.6
|
0.9
|
0.4
|
0.6
|
|
0 to 400
|
0.3
|
0.4
|
0.3
|
0.4
|
|
B
|
600 to 800
|
1.3
|
2.0
|
1.0
|
1.5
|
|
800 to 1000
|
1.0
|
1.5
|
0.8
|
1.2
|
|
1000 to 1820
|
0.9
|
1.3
|
0.8
|
1.2
|
|
R
|
-20 to 0
|
2.3
|
2.8
|
1.2
|
1.8
|
|
0 to 100
|
1.5
|
2.2
|
1.1
|
1.7
|
|
100 to 1767
|
1.0
|
1.5
|
0.9
|
1.4
|
|
S
|
-20 to 0
|
2.3
|
2.8
|
1.2
|
1.8
|
|
0 to 200
|
1.5
|
2.1
|
1.1
|
1.7
|
|
200 to 1400
|
0.9
|
1.4
|
0.9
|
1.4
|
|
1400 to 1767
|
1.1
|
1.7
|
1.0
|
1.5
|
|
C
|
0 to 800
|
0.6
|
0.9
|
0.6
|
0.9
|
|
800 to 1200
|
0.8
|
1.2
|
0.7
|
1.0
|
|
1200 to 1800
|
1.1
|
1.6
|
0.9
|
1.4
|
|
1800 to 2316
|
2.0
|
3.0
|
1.3
|
2.0
|
|
L
|
-200 to -100
|
0.6
|
0.9
|
0.3
|
0.4
|
|
-100 to 800
|
0.3
|
0.4
|
0.2
|
0.3
|
|
800 to 900
|
0.5
|
0.8
|
0.2
|
0.3
|
|
U
|
-200 to 0
|
0.6
|
0.9
|
0.4
|
0.6
|
| 0 to 600
|
0.3
|
0.4
|
0.3
|
0.4
|
|
BP
|
0 to 1000
|
1.0
|
1.5
|
0.4
|
0.6
|
|
1000 to 2000
|
1.6
|
2.4
|
0.6
|
0.9
|
|
2000 to 2500
|
2.0
|
3.0
|
0.8
|
1.2
|
|
XK
|
-200 to 300
|
0.2
|
0.3
|
0.2
|
0.5
|
| 300 to 800
|
0.4
|
0.6
|
0.3
|
0.6
|
|
|


|
|